| Item |
Parameter |
Notes |
| Appearance |
 |
(W × D × H) – [mm]: 520 × 440 × 550 |
| System Pump |
Flow Rate Range: 0–25 mL/min
Maximum Pressure: 30 MPa / 4200 psi
Flow Repeatability: 0.25–25 mL/min, <3 MPa, 0.8–2 cP
Flow Accuracy: ±1%
Flow Precision: RSD <0.5%
Gradient Accuracy: ±0.6%
Interface: Ethernet communication
Pressure Sensor: 0–20 MPa, accuracy ±0.025 MPa
Cleaning: Automatic flushing |
Dual-pump gradient system
25 mL/min per pump
Supports constant pressure mode and pressure control mode |
| Mixer |
Dynamic mixer: standard 1.4 mL, optional 5 mL or 10 mL |
Standard 1.4 mL |
| UV Detector |
Single-wavelength LED light source, Abs: ±6 AU |
Single wavelength 280 nm
Optional: 190–500 nm, 3 wavelengths
Optional: 190–850 nm, 3 wavelengths
Optional: 190–850 nm, 4 wavelengths |
| Conductivity / Temperature Monitor |
Conductivity: 0–999 mS/cm
Resolution: 0.001 mS/cm
Accuracy: ±1.0%
Automatic temperature compensation |
Standard |
| pH Valve |
pH Range: 0–14
Resolution: 0.01
Accuracy: ±0.05
7-position, 5-port valve
pH electrode and restrictor can be online or bypassed |
1: Restrictor → pH
2: Restrictor
3: pH
4: Bypass
5: Restrictor |
| Communication Interface |
TCP/IP, 100 Mbps Ethernet |
Standard |
| Accessory Kit |
Tubing, fittings and accessories |
Standard |
| Chromatography Workstation Software |
Chinese / English user interface, including instrument control, method editing,
data processing, report printing, user and result database, encrypted database
management, peak integration, quantitative analysis, audit trail, scouting,
channel scouting, method programming and 8-channel parallel operation. |
Standard |
| Pump A Inlet Valve |
Pump A inlet valve, maximum pressure 10 MPa, optional 3 / 7 / 13 channels |
12 channels + 1 common port high-pressure rotary valve (13 channels) |
| Pump B Inlet Valve |
Pump B inlet valve, maximum pressure 10 MPa, optional 3 / 7 / 13 channels |
12 channels + 1 common port high-pressure rotary valve (13 channels) |
| Sample Pump Inlet Valve |
Sample pump inlet valve, maximum pressure 10 MPa, optional 3 / 7 / 13 channels |
6 channels + 1 common port high-pressure rotary valve (7 channels) |
| Air Sensor, Inlet A |
External ultrasonic air sensor |
Standard, optional built-in ultrasonic sensor |
| Air Sensor, Inlet B |
External ultrasonic air sensor |
Standard, optional built-in ultrasonic sensor |
| Air Sensor, Sample Pump Inlet |
External ultrasonic air sensor |
Standard, optional built-in ultrasonic sensor |
| Injection Valve |
8-position, 6-port high-pressure rotary valve, pressure rating 30 MPa |
1: Manual Load
2: Inject
3: System Pump Waste
4: Sample Pump Load
5: Direct Inject
6: Sample Pump Waste |
| V Valve |
1: 1–2
2: 3–4
3: 1–3
4: 2–4
Pressure rating: 30 MPa |
None |
| Column Valve |
Configurable, pressure rating 30 MPa
Optional: 2-column valve / 2-column valve with reverse flow / 6-column valve with reverse flow |
6-column valve, 5 column flow paths, 1 bypass line |
| Column Delta Pressure Module |
Pre-column pressure, post-column pressure and delta pressure |
Standard |
| Outlet Valve |
Maximum pressure 10 MPa, optional 3 / 7 / 13 channels |
12 channels + 1 common port high-pressure rotary valve (13 channels) |
| Back Pressure Valve / Flow Restrictor |
HY-FR804 |
Standard system back pressure: 0.3 MPa |
| Fraction Collector |
1. Frac100A: Open-frame collector, compatible with standard XY100 tubes
2. Frac100B: Enclosed cabinet, 6 rack positions, compatible with 50 mL and 15 mL tubes, automatic rack recognition
3. Frac100B-1: Enclosed cabinet, 6 rack positions, compatible with 50 mL tubes, 15 mL tubes, 24-well, 48-well and 96-well plates, automatic rack recognition
4. Frac100B-2: Enclosed cabinet, 6 rack positions, compatible with 50 mL tubes, 15 mL tubes, 24-well, 48-well and 96-well plates, with UV sterilization and temperature / humidity monitoring, automatic rack recognition
5. Frac100C: Enclosed cabinet, 6 rack positions, compatible with 50 mL tubes, 15 mL tubes, 24-well, 48-well and 96-well plates, with UV sterilization, temperature / humidity monitoring and TE cooling, automatic rack recognition |
Frac100B-1 |
| Sample Pump |
Flow Rate Range: 0–100 mL/min
Maximum Pressure: 10 MPa / 1400 psi
Flow Repeatability: 1–100 mL/min, <3 MPa, 0.8–2 cP
Flow Accuracy: ±1%
Flow Precision: RSD <0.5%
Gradient Accuracy: ±0.6%
Interface: Ethernet communication
Pressure Sensor: 0–20 MPa, accuracy ±0.025 MPa
Cleaning: Automatic flushing |
Standard: 100 mL/min
Optional: 150 mL/min
Optional: 25 mL/min |
| SDK & Communication Protocol |
Optimized proprietary command set. Tested development libraries are provided,
including .NET libraries, C/C++ DLLs and sample source code. |
Suitable for secondary development. |
| Human-Machine Interface |
Windows 7 / Windows 10 / Windows 11 |
Configuration: Intel i3 processor or above, memory ≥8 GB, storage ≥256 GB, display resolution 1920 × 1080 |
| Power Supply |
110–220 VAC, ±5% |
|
| Operating Temperature |
-10°C to 40°C, non-condensing |
|